Title : Validation of two automatic devices, Omron HEM-6232T and HEM-6181, for self-measurement of blood pressure at the wrist according to the ANSI/AAMI/ISO 81060-2:2013 protocol and the European Society of Hypertension International Protocol revision 2010.

Pub. Date : 2019

PMID : 30881007






1 Functional Relationships(s)
Download
Sentence
Compound Name
Protein Name
Organism
1 Results: The mean differences between the devices and mercury readings for SBP and DBP were as follows: HEM-6232T, -0.4+-6.7 mmHg and 1.6+-5.4 mmHg, respectively; HEM-6181, -0.7+-6.2 mmHg and -0.7+-5.2 mmHg, respectively, satisfying the ANSI/AAMI/ISO protocol. Mercury selenium binding protein 1 Homo sapiens