Title : Thin layer coulometric determination of nitrate in fresh waters.

Pub. Date : 2012 Sep 26

PMID : 22935372






2 Functional Relationships(s)
Download
Sentence
Compound Name
Protein Name
Organism
1 Coulometric analysis with the tubular membrane ISE showed that nitrate could be detected in the range 10-100 muM with a precision of 2.3% relative standard deviation (RSD), limit of detection of 1.1 muM and relative accuracy of 4.4% compared to a certified reference material (CRM) Lake sample. Nitrates latexin Homo sapiens
2 Coulometric analysis with the tubular membrane ISE showed that nitrate could be detected in the range 10-100 muM with a precision of 2.3% relative standard deviation (RSD), limit of detection of 1.1 muM and relative accuracy of 4.4% compared to a certified reference material (CRM) Lake sample. Nitrates latexin Homo sapiens