Title : Comparison of the new polyethersulfone high-flux membrane DIAPES HF800 with conventional high-flux membranes during on-line haemodiafiltration.

Pub. Date : 2003 Nov

PMID : 14551370






3 Functional Relationships(s)
Download
Sentence
Compound Name
Protein Name
Organism
1 The mean plasma reduction rate of beta(2)m was 77 +/- 1% for DIAPES HF800 and polysulfone whereas it was 71 +/- 1% for polyamide (P < 0.05). polysulfone P 1700 beta-2-microglobulin Homo sapiens
2 The mean beta(2)m amount removed and found in the haemodiafiltrate per session was 230 +/- 14 mg for DIAPES HF800, 186 +/- 13 mg for polysulfone and 147 +/- 13 mg for polyamide (P < 0.05 between each pair of membranes). polysulfone P 1700 beta-2-microglobulin Homo sapiens
3 CONCLUSIONS: On-line HDF has shown to achieve plasma reduction rates for beta(2)m of up to 77% for the DIAPES HF800 membrane and for polysulfone. polysulfone P 1700 beta-2-microglobulin Homo sapiens